Specification of Multifunctional Calibrator ALL300 Brand Constant
Multifunctional Calibrator ALL300 Brand Constant
Features:
- Source: DC Voltage, DC Current, resistance, simulate Transmitter, TC, RTD, frequency, pulse and switches
- Large LCD Display Simultaneously: TC/RTD and mV/t Ω, mA and mA%
- Manual Step Source and Auto
- Room Temperature Monitoring
Reference | Range | Resolution | Accuracy |
Remark
|
DC Voltage |
100mV |
-10,000 mV ~ 110,000 mV |
1 µV |
0.12 + 0.01 |
Maximum output current: 0.5 mA |
1000mV |
-100mV ~ 1100mV |
10 µV |
d >
0.12 + 0.01
Maximum output current: 2mA |
10V |
-1.0000 V ~ 11.0000 V |
0.1 mV |
0.12 + 0.01 |
Maximum output current: Ma |
DC Current |
20mA |
0000 mA ~ 22,000 mA |
1 µA |
0.02 + 0.02 |
External supply for simulate mA: 5V ~ 28V. Maximum load 1KΩ at 20mA |
Resistance |
400Ω |
0.00Ω ~ 400.00Ω |
0.01Ω |
0.02 + 0.02 |
Excitation current: ± 0.5-3mA; If ±: 0.1-0.5 add 0.1 Ω Accuracy does not include lead resistance; |
4KΩ |
0.0000 KΩ ~ 4.0000 KΩ |
0.1Ω |
0.05 + 0025 |
Excitation current: ± 0.05-0.3 mA; Does not include lead |
resistance;
40KΩ |
0.0000 KΩ ~ 40.0000 KΩ |
1Ω |
0.1 + 0.1 |
Excitation current: ± 0.01 mA; Does not include lead resistance; |
TC |
R |
0 ° C ~ 1767 ° C |
1 ° C |
0 ~ 100 ° C: 1.5 ° C 100 ~ 1767 ° C: 1.2 ° C |
B |
y Using the ITS-90 temperature scale; The accuracy does not include the error of internal temperature compensation caused by a sensor
S |
0 ° C ~ 1767 ° C |
0 ~ 100 ° C: 1.5 ° C 100 ~ 1767 ° C: 1.2 ° C |
K |
-200.0 ° C ~ 1372.0 ° C |
0.1 ° C |
-200.0 100.0 ° C ~-: 0.6 ° C -100.0 ~ 400.0 ° C: 0.5 ° C 1200.0 ~ 400.0 ° C: 0 ° |
C
1372.0 ~ 1200.0 ° C: 0.9 ° C
E |
-200.0 ° C ~ 1000.0 ° C |
-200.0 100.0 ° C ~-: 0.6 ° C 100.0-600.0 ° C ~: 0.5 ° C 1000.0 ~ 600.0 ° C: 0.4 ° C |
J |
-200.0 ° C ~ 1200.0 ° C |
-200.0 100.0 ° C ~-: 0.6 ° C -100.0 ~ 800.0 ° c: 0.3 ° C 800.0 ~ 1200.0 ° C: 0.7 ° C |
T |
-250.0 ° C ~ 400.0 ° C |
TD >
-200.0 ~ 400.0 ° C: 0.6 ° C
N |
-200.0 ° C ~ 1300.0 ° C |
-200.0 100.0 ° C ~-: 0.6 ° C -100.0 ~ 900.0 ° C: 0.7 ° C 900.0 ~ 1300.0 ° C: 0.8 ° C |
B |
600 ° C ~ 1820 ° C |
1 ° C |
600.0 ~ 800.0 ° C: 1.5 ° C ~ 1820.0 800.0 ° C: 1.1 ° C |
L |
-200.0 ° C ~ 900.0 ° C |
|
n = "2" >
0.1 ° C
-200.0 ~ 0.0 ° C: 0.7 ° C 0.0 ~ 900.0 ° c: 0.3 ° C |
U |
-200.0 ° C ~ 600.0 ° C |
-200.0 ~ 0.0 ° C: 0.7 ° C 600.0 ° C ~ 0.0:0.5 ° C |
RTD |
Pt100-385 |
-200.0 ° C ~ 800.0 ° C |
0.1 ° C |
-200.0 ~ 0.0 ° c: 0.3 ° C 0.0 ~ 400.0 ° C: 0.5 ° C
|
800.0 ~ 400.0 ° C: 0.8 ° C
By using temperature scale ITS-90. Excitation current: ± 0.5 ~ ± 3mA for Pt100, Cu10, Cu50; Add 0.5 ° C when the excitation current is ± 0.1 ~ 0.5 mA. Excitation current: ± 0.05 mA ~ ± 0.3 mA for Pt200, Pt500, Pt1000; Does need include lead resistance. |
Pt200-385 |
-200.0 ° C ~ 630.0 ° C |
-200.0 100.0 ° C ~: 0.8 ° C 100.0 300.0 ° C ~: 0.9 ° C 300.0 ~ 630.0 ° c: 1.0 ° |
C
Pt500-385 |
-200.0 ° C ~ 630.0 ° C |
-200.0 100.0 ° C ~: 0.4 ° C 100.0 ~ 300.0 ° c: 0.3 ° C 300.0 ~ 630.0 ° C: 0.7 ° C |
Pt1000-385 |
-200.0 ° C ~ 630.0 ° C |
-200.0 ~ 100.0 ° C: 0.2 ° C 100.0 ~ 300.0 ° c: 0.3 ° C 300.0 ~ 630.0 ° C: 0.7 ° C |
Cu10 |
-100.0 ° C ~ 260.0 ° C |
1.8 ° C |
Cu50 |
-50.0 ° C ~ 150.0 ° C |
0.4 ° C |
FREQ |
100Hz |
1.00 65.00 ~ Hz Hz |
0.01 Hz |
± 2count |
Output voltage: + 1 ~ + 11Vp-p (zero base waveform); Ampiltude accuracy: ± (3% + 0.3 V); Maximum load: > 100KΩ; Duty cycle: |
50%
1KHz |
0100 KHz ~ 100 KHz |
1Hz |
10KHz |
1.0 KHz ~ 11.0 KHz |
0.1 KHz |
100KHz |
10KHz ~ 110KHz |
2KHz |
c. 5count |
PULSE |
100Hz |
1 ~ 100000cycles |
|
colspan = "1" rowspan = "3" >
1cyc
± 2count |
1KHz |
10KHz |
SWITCH |
100Hz |
1.00 65.00 ~ Hz Hz |
0.01 Hz |
± 2count |
FET switch Maximum open/close voltage: + 28V Maximum open/clo |