Specifications | |
Temperature Range | -30 °C to 650 °C (-22 °F to 1202 °F) |
Accuracy |
±1.0 °C or ±1.0% of reading, whichever is greater -10 °C to 0 °C: ±2.0 -30 °C to -10 °C: ±3.0 |
Response Time (95%) | Fluke 62 MAX+ <300 ms (95% of reading) |
Spectral Response 8 to 14 microns | |
Emissivity 0.10 to 1.00 | |
Optical Resolution | 12:1 (calculated at 90% energy) |
Display Resolution | 0.1 °C (0.2 °F) |
Repeatability of Readings | ±0.5% of reading or <±0.5 °C (1 °F), whichever is greater |
Power | AA battery |
Battery Life | 8 hours with laser and backlight on |
Physical Specifications | Weight 255 g (8.99 oz) |
Size (175 x 85 x 75) mm (6.88 x 3.34 x 2.95) inches | |
Operating Temperature 0 °C to 50 °C (32 °F to 122 °F) | |
Storage Temperature -20 °C to 60 °C (-4 °F to 140 °F), (without battery) | |
Operating Humidity 10 % to 90 % RH non-condensing at 30 °C (86 °F) | |
Operating Altitude 2000 meters above mean sea level | |
Storage Altitude 12,000 meters above mean sea level | |
IP Rating IP 54 per IEC 60529 | |
Drop Test 3 meters | |
Vibration and Shock IEC 68-2-6 2.5 g, 10 to 200 Hz, IEC 68-2-27, 50 g, 11 ms | |
EMC EN 61326-1:2006 EN 61326-2:2006 | |
Standards and Agency Approval | Compliance EN/IEC 61010-1: 2001 |
Laser Safety FDA and EN 60825-1 Class II |