Specification of Digital Multimeters: Fluke 113 Utility Multimeter
Digital Multimeters: Fluke 113 Utility Multimeter
True-rms Digital Multimeter designed for basic electrical tests
The Fluke 113 Digital Multimeter gives users the means to quickly and easily do basic meter set and reconnect testing. This meter is simple-to-use and has the features needed to repair most electrical problems. The Fluke 113 has significant improvements over other digital multimeters available on the market, including Fluke model 7-600. Using the Fluke VCHEK™ LoZ low impedance measurement function, users can simultaneously test for voltage or continuity. The Min/Max function lets you record signal fluctuations. The Fluke 113 can perform diode tests, and also provides both auto and manual ranging capabilities. With features such as conformance to the latest safety standards, backlight, and a much larger display that’s easier to view, this meter is a must have for every toolbox.
Features and benefits of the Fluke 113
VCHEK™ LoZ low impedance measurement function to simultaneously test for voltage or continuity
True-rms for accurate ac measurements on non-linear loads
A large backlit display allows for easy visibility in low-lit areas
Record signal fluctuations using the Min/Max function
Diode test for testing general-purpose germanium, silicon, and power diodes
Integral holster with probe holders for easy storage
Auto and Manual ranging for added user functionality
Meets Measurement Category CAT III 600 V and Category IV 300 V standards for a higher level of safety
The optional TPAK magnetic hanger allows the user hands-free flexibility
Specifications |
|
VCHEK 1,2,3 |
Range 6.000 V
60.00 V
600.0 V
Resolution 0.001 V
0.01 V
0.1 V
Accuracy DC, 45 to 500 Hz:2.0 % + 3
500 Hz to 1 kHz: 4.0 % + 3 |
Diode test 4 |
Range 2.000 V
Resolution 0.001 V
Accuracy 2.0 % + 3 |
Continuity4 |
Accuracy Beeper on < 20 Ω, off > 250 Ω; detects opens or shorts of 500 μs or longer |
Ω4 |
Range 600.0 Ω
6.000 kΩ
60.00 kΩ
Resolution 0.1 Ω
0.001 kΩ
0.01 kΩ
Accuracy 0.9 % + 2
0.9 % + 1
0.9 % + 1 |
Capacitance4 |
Range 1000 nF
10.00 μF
100.0 μF
9999 μF
Resolution 1 nF
0.01 μF
0.1 μF
1 μF
Accuracy 1.9 % + 2
1.9 % + 2
1.9 % + 2
100 μF - 1000 μF: 1.9 % + 2
> 1000 μF: 5 % + 20 |
Notes:
1 All VChek voltage ranges are specified from 60 counts to 100 % of range. Because inputs below 60 counts are not specified, it is possible and normal for this and other True-RMS meters to display non-zero readings when the test leads are disconnected from a circuit or are shorted together.
2 Crest factor of ≤ 3 at 4000 counts, decreasing linearly to 1.5 at full scale.
3 Minimum measureable input in VCHEK autorange is 3 V dc. For measurements below 3 V dc, push Range button twice to get to 6 V dc manual range.
4 After measuring voltage, a wait time of 1 minute is required to maintain accuracy of ohms, capacitance, diode test, and continuity.
Specifications |
|
Maximum voltage between any terminal and earth ground |
600 V |
Display |
3 3/4-digits, 6000 counts, updates 4/sec |
Operating temperature |
-10 °C to 50 °C (14 °F to 122 °F) |
Storage temperature |
-40 °C to 60 °C (-22 °F to 140 °F) |
Temperature coefficient |
0.1 x (specified accuracy)/°C (< 18 °C or > 28 °C) |
Operating altitude |
2,000 m |
Storage altitude |
10,000 m |
Relative humidity |
95 % to 30 °C
75 % to 40 °C
45 % to 50 °C |
Battery type |
9 Volt Alkaline, ANSI 1604A / IEC 6F22 |
Battery life |
Alkaline: 300 hours typical, without backlight |
Battery life |
1 Meter drop per IEC 61010-1-2001 |
Vibration |
Per MIL-PRF-28800 for Class 2 instrument |
Size (HxWxL) |
167.1 mm X 85.1 mm X 46.0 mm (6.58 in X 3.35 in X 1.81 in) |
Weight |
13.0 oz (404 g) |
Safety |
Complies with ANSI/ISA 82.02.01 (61010-1) 2004, CAN/CSA-C22.2 No 61010-1-04, UL 61010-1 (2004) and IEC/EN 61010-1 2nd Edition for measurement Category IV, 600 V, Pollution Degree 2, EMC EN61326-1 |
EMI regulations |
Complies with FCC Part 15, Class B |